Comparison of Depth Dose Distributions Using Cerenkov Fiber-Optic Dosimeter and Monte Carlo Simulation for HDR BrachytherapyReport as inadecuate




Comparison of Depth Dose Distributions Using Cerenkov Fiber-Optic Dosimeter and Monte Carlo Simulation for HDR Brachytherapy - Download this document for free, or read online. Document in PDF available to download.

In thisstudy, we fabricated a Cerenkov fiber-optic dosimeter CFOD without anyscintillator to measure Cerenkov radiation signals owing to gamma-rays. The relativedepth dose RDD distributions of Ir-192 HDR brachytherapy source were obtainedby using the CFOD based on a subtraction method and the RDD curve was comparedwith the simulation result of Monte Carlo N-particle extended transport codeMCNPX. Finally, we demonstrated that the CFOD can be used to measurereal-time dose information for HDR brachytherapy.

KEYWORDS

Cerenkov Fiber-Optic Dosimeter, Cerenkov Radiation, Gamma-Ray, Relative Depth Dose, HDR Brachytherapy

Cite this paper

Shin, S. , Song, Y. , Kim, M. , Kim, H. , Yoo, W. , Jang, K. and Lee, B. 2016 Comparison of Depth Dose Distributions Using Cerenkov Fiber-Optic Dosimeter and Monte Carlo Simulation for HDR Brachytherapy. Journal of Applied Mathematics and Physics, 4, 1499-1502. doi: 10.4236-jamp.2016.48157.





Author: Sang Hun Shin1, Young Beom Song2, Mingeon Kim2, Hye Jin Kim2, Wook Jae Yoo1, Kyoung Won Jang1, Bongsoo Lee2*

Source: http://www.scirp.org/