The Potential of Compact Libs System with Multi-Pulse Nd:YAG Laser for Bacteria IdentificationReport as inadecuate




The Potential of Compact Libs System with Multi-Pulse Nd:YAG Laser for Bacteria Identification - Download this document for free, or read online. Document in PDF available to download.

This work shows the potential of a very compact, portable Laser Induced-Breakdown Spectroscopy LIBS system that includes low cost and compact Nd:YAG laser, intended for use in fieldwork identification of bacteria. This device allows for the generation of 2 - 4 pulses per laser shot, with a maximum energy of 120 mJ per pulse. Each pulse has a duration of 60 ns, and the total duration of the train of pulses is 180 μs. The device dimensions including the power supply are 50 × 20 × 140 cm, which indicates that can be utilized as a portable LIBS system in field conditions. The spectra captured on two different bacteria shows the capability of identification.

KEYWORDS

Laser, LIBS, Bacteria, Spectroscopy

Cite this paper

Flores, A. , Sanchez, A. , Villarreal, A. , Sauz, F. , Cabrera, L. and Reyes, T. 2015 The Potential of Compact Libs System with Multi-Pulse Nd:YAG Laser for Bacteria Identification. Journal of Biomedical Science and Engineering, 8, 207-212. doi: 10.4236-jbise.2015.83020.





Author: A. Ponce Flores1*, A. K. Frias Sanchez2, A. Villarreal2, F. G. Rendón Sauz2, L. Ponce Cabrera2, T. Flores Reyes2

Source: http://www.scirp.org/