Optimization of Image Quality in Retrospective Respiratory-Gated Micro-CT for Quantitative Measurements of Lung Function in Free-Breathing RatsReport as inadecuate




Optimization of Image Quality in Retrospective Respiratory-Gated Micro-CT for Quantitative Measurements of Lung Function in Free-Breathing Rats - Download this document for free, or read online. Document in PDF available to download.

Objective: To optimize scan time and X-ray dose with no loss of image quality for retrospectively gated micro-CT scans of free-breathing rats. Methods: Five free-breathing rats were scanned using a dynamic micro-CT scanner over 10 continuous gantry rotations 50 seconds and entrance dose of 0.28 Gy. The in-phase projection views were selected and reconstructed, representing peak inspiration and end expiration from all 10 rotations and progressively fewer rotations. A least error method was also used to ensure that all angular positions were filled. Image quality and reproducibility for physiological measurements were compared for the two techniques. Results: The least error approach underestimated the lung volume, air content in the lung at peak inspiration, and tidal volume. Other measurements showed no differences between the projection-sorting techniques. Conclusions: Seven gantry rotations 35 seconds and 0.2 Gy dose proved to be the optimal protocol for both the in-phase images and the least error images.

KEYWORDS

Respiratory-Gating; Micro-CT; Lung Imaging; Free-Breathing Rats; Tidal Volume

Cite this paper

Ford, N. , Jeklin, A. , Yip, K. , Yohan, D. , Holdsworth, D. and Drangova, M. 2014 Optimization of Image Quality in Retrospective Respiratory-Gated Micro-CT for Quantitative Measurements of Lung Function in Free-Breathing Rats. Journal of Biomedical Science and Engineering, 7, 157-172. doi: 10.4236-jbise.2014.74020.





Author: Nancy L. Ford, Andrew Jeklin, Karen Yip, Darren Yohan, David W. Holdsworth, Maria Drangova

Source: http://www.scirp.org/