Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control PhantomReport as inadecuate




Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom - Download this document for free, or read online. Document in PDF available to download.

The Adaptive Quality Control Phantom AQCP is the computer-controlled phantom which positions and moves a radioactive source in the Field of View FOV of an imaging nuclear medicine device on a definite path to produce any spatial distribution of gamma rays to perform the QC Tests such as the Collimator Hole Angulation CHA of Single Photon Emission Computer Tomography SPECT. The collimator hole angulation for seven collimators were measured with the method by using a point source and computer-controlled cylindrical positioning. In this method the displacement of the image of a point source examined as the AQCP move point source vertically away from the collimator face. The results of the high-accuracy measurement method of CHA show that the measurement accuracy for absolute angulation errors is better than ±0.024°. The Root Mean Square RMS of CHA for LEHR, LEHS, LEUHR, MEGP-250, MEGP-300, MEGP-360 and HEPH collimators of SMV dual heads camera were measured to be 0.290°, 0.292°, 0.208°, 0.194°, 0.181°, 0.177°, 0.150°, respectively. The Root Mean Square RMS of CHA for LEGP, MEGP and HEGP collimators of GE Millennium MG were measured to be 0.154°, 0.220° and 0.202° respectively. It is to be added in this connection that the measured RMS of CHA for LEHR collimator with the distance variation from the collimator’s surface +-– 1 mm has been varied +-– 0.04 degree.

KEYWORDS

Nuclear Medicine; Adaptive Quality Control Phantom; Collimator hole Angulation

Cite this paper

H. Zeinali, E. Goodarzi, H. Ravanbakhsh, A. Sardarpour, N. Mahabadi, S. Moradkhani, F. Dolatshah and D. Rahi -Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom,- Modern Instrumentation, Vol. 1 No. 4, 2012, pp. 49-53. doi: 10.4236-mi.2012.14007.





Author: Hossein Zamani Zeinali, Ehsan Masumi Goodarzi, Hamid Ravanbakhsh, Ali Asgher Sardarpour, Narjes Abagheri Mahabadi, Soghra Moradkh

Source: http://www.scirp.org/