Abstract

SEFRE Shoulder-Elbow-Forearm Robotics Economic rehabilitation system is presented in this paper. SEFRE Rehab System is composed of a robotic manipulator and an exoskeleton, so-called Forearm Supportive Mechanism FSM. The controller of the system is developed as the Master PC consisting of five modules, that is, Intelligent Control IC, Patient Communication PC, Training with Game TG, Progress Monitoring PM, and Patient Supervision PS. These modules support a patient to exercise with SEFRE in six modes, that is, Passive, Passive Stretching, Passive Guiding, Initiating Active, Active Assisted, and Active Resisted. To validate the advantages of the system, the preclinical trial was carried out at a national rehabilitation center. Here, the implement of the system and the preclinical results are presented as the verifications of SEFRE.