Full validation of an electrothermal atomic absorption assay for zinc in hepatic tissue using a fast sample preparation procedureReport as inadecuate




Full validation of an electrothermal atomic absorption assay for zinc in hepatic tissue using a fast sample preparation procedure - Download this document for free, or read online. Document in PDF available to download.

Spectroscopy - Volume 20 2006, Issue 2, Pages 81-87



Programa de Pós-Graduação em Gastroenterologia, Faculdade Medicina, Universidade Federal do Rio Grande do Sul, Porto Alegre, Brazil

Programa de Pós-Graduação em Pediatria, Faculdade Medicina, Universidade Federal do Rio Grande do Sul, Porto Alegre, Brazil

Programa de Pós-Graduação em Ciências Farmacêuticas, Faculdade de Farmácia, Universidade Federal do Rio Grande do Sul, Porto Alegre, Brazil

Programa de Pós-Graduação em Ciências Farmacêuticas, Faculdade de Farmácia, Universidade Federal do Rio Grande do Sul, Ipiranga, 2752-703, CEP 90610-000, Porto Alegre, Brazil



Copyright © 2006 Hindawi Publishing Corporation. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

The objective of this work was to develop and fully validate an analytical assay to quantify zinc in hepatic tissue. The procedure should be as simple and fast as possible in order to avoid sample contamination. The amount of sample used should also reflect the sample size usually obtained in clinical biopsies, which are about 3–4 mg at most. The validation protocol is in accordance to international guidelines, such as ICH and FDA. The parameters evaluated were precision, accuracy, range, limit of detection and limit of quantification. The method was evaluated in the 2.0–32.0 parts per billion μg-l range. Under the described conditions intra and inter day precision of the three levels of quality controls were lower than 9.06 and 5.27, respectively, expressed as relative standard deviation RSD. The accuracy ranged from 86.35 to 114.71%. Limit of detection and limit of quantification were 0.60 and 2.0 μg-l, respectively. Fresh bovine liver samples were used in order to evaluate the clinical procedure used to collect biopsies. According to the results and experimental protocol, the method is fully validated and ready to use in clinical trials involving zinc quantitation using hepatic samples as small as 2.00 mg of dry tissue.





Author: Pedro Eduardo Fröehlich, Raquel Borges Pinto, André Castagna Wortmann, Richard Borba Magalhães, Ana Cláudia Reis Schneider, J

Source: https://www.hindawi.com/